Instrument Details:
Make : ASD Inc.
Model: FieldSpec 4 Hi-Res
Specifications:
Spectral Range |
350-2500 nm |
Spectral Resolution
|
3 nm @ 700 nm 8 nm @ 1400/2100 nm |
Spectral sampling (bandwidth) |
1.4 nm @ 350-1000 nm 1.1 nm @ 1001-2500 nm |
Scanning Time |
100 milliseconds |
Stray light specification |
VNIR 0.02%, SWIR 1 & 2 0.01% |
Wavelength reproducibility |
0.1 nm |
Wavelength accuracy |
0.5 nm |
Maximum radiance |
VNIR 2X Solar, SWIR 10X Solar |
Channels |
2151 |
Detectors
|
VNIR detector (350-1000 nm): 512 element silicon array SWIR 1 detector (1001-1800 nm): Graded Index InGaAs Photodiode, Two Stage TE Cooled SWIR 2 detector (1801-2500 nm): Graded Index InGaAs Photodiode, Two Stage TE Cooled |
Noise Equivalent Radiance
|
VNIR 1.0 X10-9 W/cm2/nm/sr @700 nm SWIR 1 1.4 X10-9 W/cm2/nm/sr @ 1400 nm SWIR 2 2.2 X10-9 W/cm2/nm/sr @ 2100 nm |
Working Principles:
The ASD FieldSpec 4 standard res. (FS4) is a transportable battery powered spectrometer. It is designed for measurements either in the course of field campaigns or in the laboratory. The spectral range of this highly resolving device covers wavelengths between 350 and 2500nm, the sampling rate is 0.2 seconds per spectrum. Three different detectors facilitate the recording of the spectra: a silicon-photodiode-array composed of 512 elements for the VNIR (350-1000nm) and thermoelectrically cooled InGaAs-photodiodes (Indium, Gallium, Arsenid) for each the SWIR1 (1000-1800nm) and the SWIR2 (1800-2500nm). With a spectral resolution varying from 3nm in the very short and 10nm in the farer wavelengths, the device records spectra based on the information of 2151 bands.
User Instructions:
Preparation (Handling the Device)-
Safe Handling of the Device and its further Utilities
Contact Us:
The Director, ICAR-NIASM
Email: director.niasm@icar.gov.in
Tel.: 02112 254057; Fax: 02112 254056
Hyperspectral Spectroradiometer Charges:
S.No. |
Usage type |
Rate / per sample (Rs) for ICAR |
Rate / per sample (Rs) for other Govt. organisations |
Rate / per sample (Rs) for private research organization |
Rate / per sample (Rs) for Industries |
1. |
Laboratory per sample scan using contact probe |
25 |
50 |
125 |
250 |
2. |
Laboratory per sample scan using bare optic fibre in dark room |
100 |
200 |
500 |
1000 |
+ GST as applicable |
|||||
There should be minimum 5 samples |
Home | Disclaimer | Privacy Policy | RTI | Contact Us | Screen Reader
Copyright © 2025 , All Rights Reserved @ ICAR-National Institute of Abiotic Stress Management, Malegaon (Kh), Baramati